This procedure describes the method of trimming the Splash-Proof
Thermapen at a single temperature point (32°F/0°C) by using a properly
made ice bath. The Thermapen should rarely need adjustment and
should only be performed if absolutely necessary.
Materials Required:
A coin to open the Thermapen battery compartment (USA nickel works
best)
Large kitchen cup (preferably 1 quart/1 liter size)
Crushed ice (enough to fill cup)
Paper clip
Determining if Your Thermapen Requires Adjustment:
1. Make a proper ice bath by following these steps:
a. Fill a large cup with crushed ice.
b. Add just enough water into the cup to fill in the gaps in ice so no ice floats off the bottom of the cup. Too
much water will cause errors.
c. Stir gently and wait approximately 1 minute.
2. Set the Thermapen display to read temperature in tenths (example: 32.1°F) of a degree.
If the display is set to read temperature in whole degrees (example: 32°F), change to tenths by following these
steps:
a. Turn the Thermapen off by folding the probe away and locking it into the probe retaining slot.
b. Using a coin with the appropriate thickness open the battery compartment.
c. Locate the switch bank directly above the batteries numbered 1 through 4.
d. Move switch 2 to the ON position by pushing the white switch forward and away from the number 2 using
the tip of an opened paper clip.
e. Replace the battery compartment and open the probe to confirm that the display now reads in tenths of a
degree (example 32.1°F).
3. Test the Thermapen display using the ice bath following these steps:
a. Turn the Thermapen on by opening the rotating probe.
b. Insert the tip of the metal probe into the ice bath about 2 inches (50mm) maximum.
c. Stir gently so the tip does not rest on any ice. While stirring observe the displayed reading after about 15
seconds.
d. Once the display is stable and no longer changes, write down the reading. You’ll need this if the
Thermapen requires adjustment.
4. Determine if an adjustment is necessary.
a.
If the reading from the ice bath test is within acceptable limits: 31.3°F and 32.7
°F (-0.C and +0.4°C); then
no adjustment is necessary.
b. If the reading is outside of acceptable limits proceed to making the trim adjustment.
Splash-Proof Super-Fast Thermapen
Calibration Procedure
ThermoWorks, Inc. 1762 W. 20 S. #100Lindon, UT 840
42 • ph: 801-756-7705 / 800-393-6434 • fax: 801-756-8948
www.thermoworks.com [email protected]
CR2032
3V
+
1 2 3 4
ON
1 2 3 4
ON
Switch bank
CR2032 battery
Trim adjust
Switch
(factory default shown)
Page view 0
1 2

Summary of Contents

Page 1 - Calibration Procedure

This procedure describes the method of trimming the Splash-Proof Thermapen at a single temperature point (32°F/0°C) by using a properly made ice bath.

Page 2

ThermoWorks, Inc. • 1762 W. 20 S. #100 • Lindon, UT 84042 • ph: 801-756-7705 / 800-393-6434 • fax: 801-756-8948 www.thermoworks.com • Info

Comments to this Manuals

No comments